(EHF) Dynamic characteristic values cannot be retained
Storage of dynamic characteristic values is possible from thefirst to the 1000th calculation. If the number of dynamiccharacteristic calculations exceeds 100... Show Detail
- No:7629
- Displayed Date:2021/12/07 10:15
- Category: Windows Software for 4830
-
(EHF) Operation during testing is different from the sett...
• When the step mode is [Auto], [Pause]steps will not be valid. • The test may pause for several seconds at a trig ger output step. • The time until a trig... Show Detail
- No:7628
- Displayed Date:2021/12/07 10:15
- Category: Windows Software for 4830
-
(EHF) A graph or list cannot be displayed
The peak and characteristic graph appears for the followingtests: • Program function test • Fatigue / Endurance test • Combination test • Frequency s... Show Detail
- No:7627
- Displayed Date:2021/12/07 10:15
- Category: Windows Software for 4830
-
(EHF) Load indication is not possible
The [Load ON]button can only be used after P.U.1 or P.U.2 isstarted. Show Detail
- No:7625
- Displayed Date:2021/12/07 10:15
- Updated DTM:2022/01/14 13:46
- Category: Windows Software for 4830
-
(EHF) The AGC function or the waveform distortion functio...
The AGC function and the waveform distortion function cannot be used simultaneously. Show Detail
- No:7623
- Displayed Date:2021/12/07 10:15
- Updated DTM:2022/01/14 13:45
- Category: Windows Software for 4830
-
(EHF) The peak value of the response waveform is differen...
The peak value of the response waveform is different from thepeak value of the set signal because correction targets thefrequency elements (amplitudes) of th... Show Detail
- No:7622
- Displayed Date:2021/12/07 10:15
- Category: Windows Software for 4830
-
(EHF) Distortion occurs in the response waveform
If the target level is too small, correction may overshoot anddisrupt the respon se waveform. Apply an actual load todetermine a suitable value. Show Detail
- No:7621
- Displayed Date:2021/12/07 10:15
- Category: Windows Software for 4830
-
(EHF) When automatic PID setting is performed, the piston...
When obtaining the test force gain for a compression orbending test , preload the te st piece. During auto tuning,I-ACTis shut off, and if the compression fo... Show Detail
- No:7620
- Displayed Date:2021/12/07 10:15
- Category: Windows Software for 4830
-
(EHF) An alarm cannot be cleared
When multiple controllers are used, simultaneous alarmoperates, and thus it may not be possible to clear the alarms ofall controllers at once. In this case, ... Show Detail
- No:7619
- Displayed Date:2021/12/07 10:15
- Category: Windows Software for 4830
-
(EHF) The entered value cannot be set
Do not change the language in the regional and languageoptions of your operating system, even if the operating systemis compatible. If the display and format... Show Detail
- No:7618
- Displayed Date:2021/12/07 10:15
- Category: Windows Software for 4830
71-80 of Total 211