(EZ-Test) Inspection of Test Force
Check wheter the test force is approximately correct or not. 1. Prepare the test force calibration jig. 2. Insert the joint for UJ into the bo ttom... Show Detail
- No:7705
- Displayed Date:2022/01/20 14:21
- Updated DTM:2022/01/20 14:35
- Category: EZ-Test Series
-
(AGS-X) Inspection of Safety Equipment
1.While pressing a jog button to move the crosshead, press the emergency stop button. 2.The crosshead should stop and not restart, even if the jog button is... Show Detail
- No:8005
- Displayed Date:2022/01/20 14:07
- Category: AGS-X Series
- Brached FAQ
-
(EZ-Test) Inspection of Safety Equipment
1.While pressing a jog button to move the crosshead, press the emergency stop button. 2.The crosshead should stop and not restart, even if the jog button is... Show Detail
- No:7703
- Displayed Date:2022/01/20 14:07
- Category: EZ-Test Series
- Brached FAQ
-
(AGS-X) Attaching Specimens (Bending Test)
1. Adjust the distance between supports. 2. Place the specimen in the center of thesupports. 3. Use jog operations to move the bendingpunch a little above ... Show Detail
- No:8003
- Displayed Date:2022/01/20 13:37
- Category: AGS-X Series
-
(EZ-Test) Attaching Specimens (Bending Test)
1. Adjust the distance between supports. 2. Place the specimen in the center of thesupports. 3. Use jog operations to move the bendingpunch a little above ... Show Detail
- No:7701
- Displayed Date:2022/01/20 13:37
- Category: EZ-Test Series
-
(AGS-X) Setting Crosshead Stroke Limits
To prevent the crosshead (or attached load cells or jigs) from colliding with anything, secureupper and lower limit collars at respective crosshead upper and... Show Detail
- No:8001
- Displayed Date:2022/01/20 13:29
- Category: AGS-X Series
-
(EZ-Test) Setting Crosshead Stroke Limits
To prevent the crosshead (or attached load cells or jigs) from colliding with anything, secureupper and lower limit collars at respective crosshead upper and... Show Detail
- No:7699
- Displayed Date:2022/01/20 13:29
- Category: EZ-Test Series
-
(AGS-X) Attaching Compression and Bending Test Jigs
1. Screw in the compression loading jig. 2.Temporarily loosen the set bolt, theninsert the upper compression plate. 3.Insert the boss on the lowercompr... Show Detail
- No:7999
- Displayed Date:2022/01/20 11:26
- Category: AGS-X Series
- Brached FAQ
-
(EZ-Test) Attaching Compression and Bending Test Jigs
« For load cell with 1kN or larger capacity» 1. Attaching the compression load jig Screwtighten the compression loadjig of type (a) or (b) (see the fi... Show Detail
- No:7697
- Displayed Date:2022/01/20 11:26
- Category: EZ-Test Series
- Brached FAQ
-
1. Attaching the Specimen 1) Insert the specimen at thecenter of the grip faces. 2) Tighten securely with bothhands. 3) Use jog operations to... Show Detail
- No:7997
- Displayed Date:2022/01/20 10:55
- Category: AGS-X Series
31-40 of Total 211