(AGX-V) Maintaining the Ball Screws
Regularly grease the ball screws to keep them lubricated. • Grease name: General grease NLGI No. 2 (Shell Alvania Grease No. 2) Part Number:339-89306-21... Show Detail
- No:8065
- Displayed Date:2022/02/01 09:52
- Category: AGX-V Series
-
(AGX-V) Maintaining the Guide Poles
For "table-top" testers, regularly greasethe guide poles. • Grease name: Molybdenum disulfidelithium-based grease Part Number: 017-27014-02 GREASE,MO-S2... Show Detail
- No:8066
- Displayed Date:2022/02/01 09:48
- Category: AGX-V Series
-
(AGX-V) Checking the Stroke Speed
1. Press the manual button on the smart controller to turn on the manual modeand record the current jog speed. 2. On the [Jog speed] setting screen, set the... Show Detail
- No:8064
- Displayed Date:2022/02/01 09:26
- Category: AGX-V Series
-
(AGX-V) Checking the Test Force
Use light weights to simply check whether the test force is almost correct. (The following values are an example.) Use a 100N weight for 1kN to 10kN load c... Show Detail
- No:8063
- Displayed Date:2022/01/31 16:37
- Category: AGX-V Series
-
(AGX-V) Configuring/Checking the Safety Devices
Set the crosshead upper limit (1) and crosshead lower limit (2) to the positions of theupper and lower limits of the crosshead movement (3). Set them to the ... Show Detail
- No:8061
- Displayed Date:2022/01/31 16:31
- Category: AGX-V Series
- Brached FAQ
-
(AGX-V) Inspection of Safety Devices
1. Set the crosshead upper limit at 2cm to 3cm above the crosshead and movethe crosshead upward with the jog up button. 2. Check that the crosshead stops wh... Show Detail
- No:8062
- Displayed Date:2022/01/31 16:28
- Category: AGX-V Series
- Brached FAQ
-
Height of the crosshead can be adjusted in the jog operation. • Pressing the manual button (2) of the smart controller enables the jog operation(manual mode... Show Detail
- No:8060
- Displayed Date:2022/01/31 15:39
- Category: AGX-V Series
- Brached FAQ
-
(AGX-V) Attaching/Replacing a Load Cell
1. Turn off the instrument. NOTE: Before replacing a load cell, turn off the power or disconnect the CAL connectorfrom the smart controller according to th... Show Detail
- No:8047
- Displayed Date:2022/01/31 14:36
- Updated DTM:2022/02/01 14:11
- Category: AGX-V Series
-
(AGX-V) Calibration of a Load Cell (E-CAL)
From the LCD touch panel of the smart controller, execute E-CAL (electroniccalibration). « When starting up» 1. Follow the message displayedafter war... Show Detail
- No:8046
- Displayed Date:2022/01/31 14:28
- Category: AGX-V Series
-
Perform the following in case of abnormality that the crosshead does not stop or thereis an odor of burning or in case of power failure. Before starting to ... Show Detail
- No:8043
- Displayed Date:2022/01/27 15:10
- Category: AGX-V Series