FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis > Inductively Coupled Plasma Emission Spectroscopy(ICP) > Multi type(ICPE)
Keyword Search
Frequently Referenced FAQ
FAQ within [ Multi type(ICPE) ]
-
(ICPE) Mounting the Nozzle (Autosampler)
1. Insert the nozzle (Parts number: 206-50290-91) into the nozzle hole located on the arm. 2. Lower the nozzle tip by 103 mm from the bottom of the arm an... Show Detail
- No:6740
- Date of Issue:2021/08/03 10:21
- Brached FAQ
-
(ICPE) Replacing the Peristaltic Pump Tubing (Autosampler)
If the flow rate of the peristaltic pump used to supply rinse solution decreases, replace the tubing in the peristaltic pump according to the following proce... Show Detail
- No:6741
- Date of Issue:2021/08/03 10:51
- Brached FAQ
-
(ICPE) Filter Cleaning and Replacement (Autosampler)
If the filter is clogged, remove the cover on the left side of the ASC-9800 and clean the filter. If the filter has deteriorated, replace it with a new filt... Show Detail
- No:6742
- Date of Issue:2021/08/03 10:56
- Brached FAQ
21-23 of Total 23