(ICPE) Stopping the ICPE in Daily Use
1. Turn off the plasma. 2. Clean the inside of the stand and the glassware. 3. Waste liquid is treated. 4. After stopping the vacuum pump,... Show Detail
- No:6732
- Date of Issue:2021/07/30 09:13
- Last Updated:2021/10/04 09:46
- Brached FAQ
-
(ICPE) Leak Valve Operation (After vacuum pump is stopped)
NOTE: When stopping the vacuum pump, release the molecular sieve case toatmospheereic pressure. 1.After stopping the vacuum pump, open the leak valve.... Show Detail
- No:6731
- Date of Issue:2021/07/30 09:07
-
(ICPE) Starting the ICPE in Daily Use
1. Supply cooling water to the ICPE. If mains water is used, open the main valve of the mains water supply. If a cooling water circulation system is used... Show Detail
- No:6730
- Date of Issue:2021/07/30 09:03
- Last Updated:2023/12/27 15:49
-
(ICPE) Cleaning the Axial-View Observation Orifice Assembly
There are ICPE-9800 Series Standard: 211-87758-42 Orifice L2 ASSY (formini torch ICPE-9800 Series Option: 211-88571-42 Orifice S2 ASSY (for optionalpla... Show Detail
- No:6729
- Date of Issue:2021/07/29 16:33
- Last Updated:2021/10/04 10:16
-
(ICPE) Cleaning and Replacement of Reflect Mirror in Cool...
1. Loosen the mirror holder's set screws and remove the mirror holder from thejacket. 2. When cleaning the mirror, leave it on the mirror holder and c... Show Detail
- No:6728
- Date of Issue:2021/07/29 16:16
- Last Updated:2021/10/04 09:47
- Brached FAQ
-
(ICPE) Attaching and Removing the Cooling Jacket
1. Remove the two cooling water tubes. 2. Remove the argon gas purge tube. 3. Remove the two attachment screws. NOTE: When the cool... Show Detail
- No:6727
- Date of Issue:2021/07/29 16:00
- Last Updated:2021/10/04 09:48
- Brached FAQ
-
(ICPE) Replacing the Peristaltic Pump Tube
Since the sample pressure feed tube is constrained by the rotor and guide, it is subjectto plastic deformation and eventually samples will no longer feed thr... Show Detail
- No:6726
- Date of Issue:2021/07/29 14:31
- Brached FAQ
-
(ICPE) Replacing and Inspecting Glassware, Sample Introdu...
WARNING: There is a risk of burn injury. Do NOT touch the plasma standinterior for 3 minutes after the plasma is extinguished. CAUTION: Glass parts break ... Show Detail
- No:6724
- Date of Issue:2021/07/29 14:03
- Brached FAQ
-
(ICPE) Maintenance for Vacuum Pump
The molecular sieve in the vacuum pump at the back of the ICPE-9800 can be replacedwhile maintaining a vacuum state in the spectrometer. Replace the molecul... Show Detail
- No:6721
- Date of Issue:2021/07/28 10:55
- Last Updated:2021/10/04 09:49
- Brached FAQ
-
(ICPE) Cleaning the Fan Filters
The fan filter on the left side of the main unit must be cleaned on a regular basis. Clean this filter once every three months. Parts Number: 211-88556 F... Show Detail
- No:6720
- Date of Issue:2021/07/28 10:19
- Last Updated:2021/10/04 09:49
- Brached FAQ