FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis > Inductively Coupled Plasma Emission Spectroscopy(ICP) > Multi type(ICPE)
Keyword Search
Frequently Referenced FAQ
FAQ within [ Multi type(ICPE) ]
-
(ICPE) Starting the ICPE in Daily Use
1. Supply cooling water to the ICPE. If mains water is used, open the main valve of the mains water supply. If a cooling water circulation system is used... Show Detail
- No:6730
- Date of Issue:2021/07/30 09:03
- Last Updated:2023/12/27 15:49
-
(ICPE) Leak Valve Operation (After vacuum pump is stopped)
NOTE: When stopping the vacuum pump, release the molecular sieve case toatmospheereic pressure. 1.After stopping the vacuum pump, open the leak valve.... Show Detail
- No:6731
- Date of Issue:2021/07/30 09:07
-
(ICPE) Stopping the ICPE in Daily Use
1. Turn off the plasma. 2. Clean the inside of the stand and the glassware. 3. Waste liquid is treated. 4. After stopping the vacuum pump,... Show Detail
- No:6732
- Date of Issue:2021/07/30 09:13
- Last Updated:2021/10/04 09:46
- Brached FAQ
-
1. Recovery after Emergency Stop After stopping the ICPEby turning OFF the power switch, perform the followingrecovery procedure. 1) Turn OFF the comp... Show Detail
- No:6733
- Date of Issue:2021/07/30 09:19
-
(ICPE) Cleaning the Entrance Window
WARNING:There is a risk of burn injury.Do NOT start work inside the plasma stand for 3 minutes afterthe plasma is extinguished. NOTE: If it takes much... Show Detail
- No:6734
- Date of Issue:2021/07/30 09:52
- Brached FAQ
-
(ICPE) Cleaning/Replacing the Purge Tube(ICPE-9820)
When a high concentration sample is analyzed, some of the sample may adhere to thepurge tube. If the tube becomes heavily contaminated, remove the purge tube... Show Detail
- No:6735
- Date of Issue:2021/07/30 09:57
-
(ICPE) The spectral intensity dropped significantly
• Deposits adhered to the tip of the nebulizer. • Deterioration of nebulizer • Deterioration of sample introduction tube • Bubbles entered the sample intr... Show Detail
- No:6736
- Date of Issue:2021/07/30 10:09
-
(ICPE) The spectral intensity is unstable
• The ICPEinstallation location does not meetrequirements related to installation location or facilities. • One of the causes listed for problem "The spectr... Show Detail
- No:6737
- Date of Issue:2021/07/30 10:11
-
(ICPE) Condensation occurs on the RF coil or the cooling ...
• Wipe off the water droplets with gauze or some otherappropriate material and ignite the plasma. • Checks that the installation location meets the installa... Show Detail
- No:6738
- Date of Issue:2021/07/30 10:11
- Last Updated:2021/10/04 10:07
-
(ICPE) The spectral intensity for elements is high even w...
• The blank liquid is dirty. • The torch is dirty. • The chamber is dirty. Click here for "Replacing and Inspecting Glassware, Sample Introduction Tub... Show Detail
- No:6739
- Date of Issue:2021/07/30 10:11
11-20 of Total 27