Frequently Referenced FAQ

FAQ within [ Sequential type(ICPS) ]

11-18 of Total 18

2 / 2page
  • (ICPS) Waste Liquid Tank

    Check the tank periodically to make sure the following. ・ The waste liquid is not overflow. ・ The drain tube does not come in contact with the liquid surfa... Show Detail

    • No:6772
    • Date of Issue:2021/08/18 15:08
    • Last Updated:2021/09/13 09:15
  • (ICPS) Cooling Water Tank

    Approximately 10L of pure water is used as the coil cooling water. Replace the water withnew one when fur forms in it. WARNIGN:You may get an electric... Show Detail

    • No:6771
    • Date of Issue:2021/08/18 15:08
  • (ICPS) Cleaning of Orifice for Axial Observation(ICPS-7510)

    NOTE: Cleaning the orifice prevents contamination of a sample and abnormaldischarge. WARNING:Never touch the inside of the stand for three minutes after e... Show Detail

    • No:6769
    • Date of Issue:2021/08/18 15:05
    • Last Updated:2022/01/17 15:47
  • (ICPS) Maintenance in the plasma stand(ICPS-7510)

    1.Cleaning part (1) Sheet copper connected with RF coil and the connecting part (2) Connecting part of this sheet copper and matching-Box.And the circumf... Show Detail

    • No:6770
    • Date of Issue:2021/08/18 15:04
  • (ICPS) Cleaning of Folding Mirror for Axial Observation(I...

    WARNING:You may get a burn. Before starting the procedures below, turn OFF theplasma and wait until the water cooled jacket is cooled sufficiently (at least3... Show Detail

    • No:6768
    • Date of Issue:2021/08/18 14:50
  • (ICPS) Operation Check of Safety Device

    WARNING: You may a burn. Don't touch the exhaust pipe on a plasma stand, or don't look in. 1) Argon gas pressure switch If you stop supplying argo... Show Detail

    • No:6766
    • Date of Issue:2021/08/18 12:48
    • Brached FAQ
  • (ICPS) Check and Replacement of Glassware

    1) Holding the chamber, remove the ball joint clip. 2) Loosen the clamp of torch stand, pull out the torch downward and remove the tube connec... Show Detail

    • No:6767
    • Date of Issue:2021/07/29 14:03
    • Brached FAQ
  • (ICPS) Maintenance of Vacuum Pump

    The molecular sieve can be replaced by a new one with the spectrometerkept at a vacuum status. Parts number: 239-07003-08 Molecular siev Parts numb... Show Detail

    • No:6775
    • Date of Issue:2021/07/28 10:55
    • Brached FAQ

11-18 of Total 18

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.