FAQ - Frequently Asked Questions
- Products Search > Continuous Monitoring Analysis
Keyword Search
Frequently Referenced FAQ
FAQ within [ Continuous Monitoring Analysis ]
-
(GAS) The indication of the connected analyzer is too low...
Improper installation of the filter element. Reinstall the filter element correctly. When this is done, check the O-ring for any damage. Click here for... Show Detail
- No:6827
- Date of Issue:2021/08/20 13:39
-
(GAS) The water drops from the gas outlet (CFP-8000)
Disabled cooler, drain pump failure. Replace the cunsumable parts. Click here for "Maintenance (Replacement of consumables)". Show Detail
- No:6826
- Date of Issue:2021/08/20 13:39
-
(GAS) The gas flows only at 1L/min (CFP-8000)
Clogged filter. Frozen cooler. Too much throttled needle valve. Mixed dust into the pump. Replace the consumable parts. Click here for "Maintenance (Re... Show Detail
- No:6825
- Date of Issue:2021/08/20 13:39
-
(GAS) The gas dose not flow at all (CFP-8000)
Disconnected piping. Broken pump. Check the piping after the pump for disconnection. About the pump manintenance, click here for "Maintenance (Replacem... Show Detail
- No:6824
- Date of Issue:2021/08/20 13:39
-
(GAS) Gas does not flow into the flow meter (CGT-7100)
• There is a problem with thepump.→ Replace the pump diaphragm orreplace thepump completely. Contact a Shimadzu sales or service representative. • There... Show Detail
- No:6823
- Date of Issue:2021/08/20 11:25
-
(GAS) The indicator drifts (CGT-7100)
• There is alarge change inambient temperature.→ Select a location with as little possible changein temperature. • The inner surface of the cell isdirty... Show Detail
- No:6822
- Date of Issue:2021/08/20 11:25
-
(GAS) The indicator values fluctuate or exhibit noise (CG...
• The ground is not complete,→ Connect a ground having a resistance of no more than 100 ohms. • There is a large source ofvibration in the vicinity.→ Re... Show Detail
- No:6821
- Date of Issue:2021/08/20 11:25
-
(GAS) O2 sensor's response becomes slow (NOA-7100)
• O2 sensor may be deteriorated. Contact a Shimadzu sales or service representative. Show Detail
- No:6820
- Date of Issue:2021/08/20 11:25
-
(GAS) Response is slow when performing span calibration o...
• Any leakage from the standard gas inlet?→Attach a hose clamp to the standard gas inlet to prevent leakage. • Any leakage from the piping inside the in... Show Detail
- No:6818
- Date of Issue:2021/08/20 11:25
- Last Updated:2021/08/23 12:49
-
(GAS) Ozone odor is evident (NOA-7100)
Has temperature of ozone deactivator risen. (The ozone deactivator is normal if it is warm to the touch.) Contact a Shimadzu sales or service representati... Show Detail
- No:6817
- Date of Issue:2021/08/20 11:25
1-10 of Total 17