FAQ within [ Continuous Monitoring Analysis ]
-
The contrast on the LCD screen may be adjusted using the procedure described below. This procedurecan be performed on all screens except the initial scree... Show Detail
- No:6808
- Date of Issue:2021/08/19 15:47
- Brached FAQ
-
Prepare a zero gas and a span gas suitable for the measurement range. *1 Press [CAL] to change the screen to [Calibration]. Set the span gas concentration, ... Show Detail
- No:6809
- Date of Issue:2021/08/19 15:47
- Brached FAQ
-
(GAS) Maintenance (Replacement of consumables)
Parts number:638-52383NH3Absorber(6pc) (1pc used) (Replaced about every month) Parts number:638-52352-02NH3Absorber(1pc) (1pc used) (Replaced about every ... Show Detail
- No:6810
- Date of Issue:2021/08/20 11:16
- Brached FAQ
-
(GAS) Gas leak checking (CGT-7100)
1. S witch the select valve to SMPL without any pipe connected to SMPL. 2. Turn on the power switch and press the MEAS key. Then adjust the needle val... Show Detail
- No:6812
- Date of Issue:2021/08/20 11:19
-
(GAS) No measurement signal (NOA-7100)
• [MEAS] key lamp on?→ Press [MEAS] key for measurement state. • Are sample and air flowing?→Reset gas flow rate. If not resolved, contact a Shimad... Show Detail
- No:6813
- Date of Issue:2021/08/20 11:25
-
(GAS) Large zero/span drift (NOA-7100)
• Is ambient temperature within permitted range?→ Change installation site. (Or, is there a change in ambient tempera ture of 10ºC or greater?) • Are samp... Show Detail
- No:6815
- Date of Issue:2021/08/20 11:25
-
(GAS) NOx signal error is large (NOA-7100)
• Is O3 flowing properly?→ Reset flow rate. • Is O3 completely dehumidified?→ Replace silica gel. If not resolved, contact a Shimadzu sales or serv... Show Detail
- No:6816
- Date of Issue:2021/08/20 11:25
-
(GAS) Ozone odor is evident (NOA-7100)
Has temperature of ozone deactivator risen. (The ozone deactivator is normal if it is warm to the touch.) Contact a Shimadzu sales or service representati... Show Detail
- No:6817
- Date of Issue:2021/08/20 11:25
-
(GAS) Response is slow when performing span calibration o...
• Any leakage from the standard gas inlet?→Attach a hose clamp to the standard gas inlet to prevent leakage. • Any leakage from the piping inside the in... Show Detail
- No:6818
- Date of Issue:2021/08/20 11:25
- Last Updated:2021/08/23 12:49
-
(GAS) O2 sensor's response becomes slow (NOA-7100)
• O2 sensor may be deteriorated. Contact a Shimadzu sales or service representative. Show Detail
- No:6820
- Date of Issue:2021/08/20 11:25