• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Optical Emission Spectrometer (PDA) ]

1-10 of Total 16

1 / 2page
  • (PDA) Checking the Water Level in the Argon Gas Bubbler

    Parts Number:211-51930 AR GAS BUBBLER (With parts and bottle) Parts Number: 038-00236 BOTTLE,PE WIDE MOUTH 1LT The argon gas bubbler is attached to... Show Detail

    • No:6513
    • Date of Issue:2021/06/25 12:57
    • Last Updated:2021/07/30 13:25
  • (PDA) Replacing the Pump Oil

    Parts Number: 017-30166-01 Pump oil, R-2 1L (for GHD-101C) Parts Number: 211-78236-91 OIL MIST TRAP KIT (Option) Parts Number: 040-82714-51 ELEMENT,OMS... Show Detail

    • No:6514
    • Date of Issue:2021/06/25 13:24
    • Brached FAQ
  • (PDA) Manual scanning (Check of entrance slit)

    Changes in the environment over the course of timecausing temperature changes inside the system maygive rise to deviations in the peak position of thespectru... Show Detail

    • No:6515
    • Date of Issue:2021/06/25 13:51
    • Brached FAQ
  • (PDA) Lamp Test

    Lamp test is a way to check behavior of the read-outsystem. Measure the intensities using the lamp test tocheck the stabilization of the whole read-out syste... Show Detail

    • No:6519
    • Date of Issue:2021/06/25 14:04
    • Brached FAQ
  • (PDA) File Maintenance

    • Making a backup copy This section describes how to make a backup copy of theanalysis information and results to removable media. 1.Open the [File Mainte... Show Detail

    • No:6524
    • Date of Issue:2021/06/28 10:10
    • Brached FAQ
  • (PDA) Replacing the Argon Gas Cylinder

    When the displayed value for the regulator (1), whichis attached to the argon gas cylinder (hereinafter, thecylinder), becomes 2 MPa or less, replace the cyl... Show Detail

    • No:6525
    • Date of Issue:2021/06/28 10:40
  • (PDA) Replacing the Recalibration Sample

    1. Print out the [Analytical Group Information] –[Recalibration] windows for the analytical group tobe changed. 2.Clean the emission table. 3.C... Show Detail

    • No:6526
    • Date of Issue:2021/06/28 11:01
    • Brached FAQ
  • (PDA) Emergency stop and recovering (PDA-7000)

    CAUTION: If any problem should occur on the instrument, stop the instrumentby emergency stop and immediately remove the cause of theproblem. 1. Emerge... Show Detail

    • No:6527
    • Date of Issue:2021/06/28 11:24
    • Last Updated:2021/08/02 15:38
  • (PDA) Daily Startup / Shutdown

    1. Turn ON the power to the printer and monitor. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. Confirm that the ... Show Detail

    • No:6528
    • Date of Issue:2021/06/28 12:54
    • Brached FAQ
  • (PDA) Startup after a long period shut down

    1. Confirm that the room temperature is in the range of 15 °C to 30 °C. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. ... Show Detail

    • No:6529
    • Date of Issue:2021/06/28 14:42
    • Brached FAQ

1-10 of Total 16

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.