FAQ - Frequently Asked Questions
Keyword Search
Frequently Referenced FAQ
FAQ within [ Error/Warning ]
-
Cooling fan error [Error Code:35]
The performance decreases in this situation.Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representativ... Show Detail
- No:5894
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:37
-
CPU register error [Error Code:38]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5893
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:36
-
PRG current error [Error Code:39]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5892
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:34
-
Power controller board error [Error Code:40]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5891
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:34
-
Flow# DC12V was out of range [Error Code:60 63 66 69 72 75]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5890
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:32
-
Flow# DC5V was out of range [Error Code:61 64 67 70 73 76]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5889
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:31
-
Flow# DC24V was out of range [Error Code:59 62 65 68 71 74]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5888
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:29
-
CARx calc. flow out of range [Error Code:5009 5010 5011 5...
Set the split ratio, column flow, and purge flow so that the total flow is within the AFC control range. Total flow is calculated with the following equat... Show Detail
- No:5887
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:25
-
CARx calc. prss out of range [Error Code:5006 5007 5008 5...
Set the linear velocity and column flow so that the column inlet pressure is within the AFC control range. TechnicalDocument_AFCnAPC-Control-Ran... Show Detail
- No:5886
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:22
-
Program time error [Error Code:4302]
Do not set the total time longer than 9999.99 min. Show Detail
- No:5885
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:13
71-80 of Total 221