FAQ - Frequently Asked Questions
Keyword Search
Frequently Referenced FAQ
FAQ within [ Error/Warning ]
-
DetAPCn DCG set point not reached[Error Code:2144 2145 21...
・Check that the column is connected. ・Check that the gas main valve is open. ・Check that the gas supply pressure is within the range below. ・C... Show Detail
- No:5981
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:38
-
DetAPCn air valve not completely closed[Error Code:2059 2...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5982
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:41
-
APCy valve not completely closed[Error Code:2069 2070 207...
・Check that the tubing connection destination of the AUX-APC is not blinded. If it is blinded, remove the blind to open the line to atmosphere. ・If the erro... Show Detail
- No:5983
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:45
-
Transmission time out[Error Code:4001]
For Ethernet connection, replace the LAN cable. Show Detail
- No:5984
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:50
-
Message is not accepted[Error Code:4003]
For Ethernet connection, replace the LAN cable. Show Detail
- No:5985
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:52
-
Command is invalid[Error Code:4005]
For Ethernet connection, replace the LAN cable. Show Detail
- No:5986
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:53
-
Param by TRS is out of range[Error Code:4006]
For Ethernet connection, replace the LAN cable. Show Detail
- No:5987
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:55
-
Transmission port is shut down[Error Code:4007]
For Ethernet connection, replace the LAN cable. Show Detail
- No:5988
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:56
-
Transmission file error[Error Code:4008]
For Ethernet connection, replace the LAN cable. Show Detail
- No:5989
- Date of Issue:2021/05/18 15:35
- Last Updated:2021/06/04 14:57
-
Gas Selector# sub set point not reached [Error Code:2350 ...
Check whether gas is supplied and whether gas is not leaking from connections. Gas leaks may be occurring in several locations (other than displayed site). Show Detail
- No:5921
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/12 01:29
61-70 of Total 221