FAQ - Frequently Asked Questions
Keyword Search
Frequently Referenced FAQ
FAQ within [ Error/Warning ]
-
DAMCn makeup was overpressure[Error Code:2288 2289 2290 2...
Check the supply gas pressure. If the supply pressure is correct, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sale... Show Detail
- No:6041
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 09:10
-
CARn purge control irregularity[Error Code:2091 2092 2093...
Check that the supply pressure to each unit is within the range below. Check that the supply pressure is sufficient for the gas of each unit to flow. Che... Show Detail
- No:6042
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 09:08
-
DetAPCn makeup control irregularity[Error Code:2100 2101 ...
Check that the supply pressure to each unit is within the range below. Check that the supply pressure is sufficient for the gas of each unit to flow. Che... Show Detail
- No:6043
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 09:05
-
APCy control irregularity[Error Code:2112 2113 2114 2115 ...
Check that the supply pressure to each unit is within the range below. Check that the gas supply pressure is stable. If the error appears repeatedly ... Show Detail
- No:6044
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 08:59
-
AOCm CH2 command error[Error Code:6007 6029]
Set the auto sampler use and sub injector use according to the actual device configuration. If the error occurs again after taking the measures, turn ... Show Detail
- No:6045
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 11:08
-
AOCm rack error[Error Code:6001 6023]
[In case of AOC-30i] Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. [In case o... Show Detail
- No:6046
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 11:23
-
AOCm syringe error[Error Code:6002 6024]
[In case of AOC-30i] • Check that the auto injector is mounted correctly and not at an angle. • Replace the syringe. If the error appears even after... Show Detail
- No:6047
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 11:24
-
AOCm plunger error[Error Code:6003 6025]
• If the plunger is bent, replace the syringe. • If the plunger is not working properly, clean the plunger or replace the syringe. • Revise the wash count ... Show Detail
- No:6048
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 14:15
-
AOCm can not start[Error Code:6004 6026]
Close the door, remove the cause, and press the [RESET] button. If the error occurs again after taking the measures, turn off the main power of the instrume... Show Detail
- No:6049
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 14:27
-
AOCm sample vial is not set[Error Code:6008 6030]
[In case of AOC-30i] • Check whether vials conforming to the equipment are being used. • Check whether the operating environment is appropriate... Show Detail
- No:6050
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 15:38
21-30 of Total 221