FAQ within [ Error/Warning ]
-
Heat is escaping[Error Code:1001]
・When oven door opens, close the door and press [Reset]. ・When the insulation of the oven has a hole, turn off the main power of the instrument and close th... Show Detail
- No:5962
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/11 15:01
-
COL CRG is out of control[Error Code:1034]
・When oven door opens, close the door and press [Reset]. ・When coolant becomes empty, turn off the main power of the instrument and replace coolant. ・If th... Show Detail
- No:5963
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 10:11
-
DET#n set temperature is low[Error Code:1049 1050 1051 1052]
・To prevent detector contamination, set a detector temperature higher than the column oven temperature. ・When setting a detector temperature lower than the ... Show Detail
- No:5964
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 10:26
-
Oven door is open[Error Code:1061]
Close column oven door and press [Reset]. If the error appears even with the oven door closed, turn off the main power of the instrument, stop gas supply, a... Show Detail
- No:5965
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:19
-
DetAPCn makeup set point not reached[Error Code:2014 2017...
・Check that the gas main valve is open. ・Check that the gas supply pressure is within the range below. ・Check that the supply pressure is sufficien... Show Detail
- No:5966
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:26
-
DetAPCn hydrogen set point not reached[Error Code:2015 20...
・Check that the gas main valve is open. ・Check that the gas supply pressure is within the range below. ・Check that the supply pressure is sufficient ... Show Detail
- No:5967
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:32
-
DetAPCn air set point not reached[Error Code:2016 2019 20...
・Check that the gas main valve is open. ・Check that the gas supply pressure is within the range below. ・Check that the supply pressure is sufficient ... Show Detail
- No:5968
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:39
-
APCy set point not reached[Error Code:2026 2027 2028 2029...
・Check that the gas main valve is open. ・Check that the gas supply pressure is higher than the set pressure. ・Check that no tubing connections are loose an... Show Detail
- No:5969
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:44
-
SCD hydrogen was overpressure[Error Code:6164]
・Check that the gas supply pressure is within the range below. ・If the gas supply pressure is appropriate, turn off the main power of the in... Show Detail
- No:5970
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:49
-
DET#1 2 3 4 flame is out[Error Code:4109 4110 4111 4112]
Check that the column is connected. Check also that there is no leak at column connections. Check that the column insertion length is appropriate. Ca... Show Detail
- No:5849
- Date of Issue:2021/05/18 15:38
- Brached FAQ