FAQ - Frequently Asked Questions
Keyword Search
Frequently Referenced FAQ
FAQ within [ Error/Warning ]
-
AOC-20s can not start[Error Code:6017]
Press the auto injector [RESET] button. Show Detail
- No:6061
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 15:28
-
AOC-20s up/down error[Error Code:6016]
Check and perform the following. • Check whether the initialization process at power-on can be completed. • Redo the teaching using the vial to be used for... Show Detail
- No:6060
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:20
-
AOC-20s expansion and control error[Error Code:6015]
Redo the teaching using the vial to be used for actual analysis. [In case of AOC-30i] AOC-30i_AOC-20sU_SamplerTeaching.pdf [In case of AOC-20i] ... Show Detail
- No:6059
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:19
-
AOC-20s rotating error[Error Code:6014]
Redo the teaching using thea vial to be used for actual analysis. [In case of AOC-30i] AOC-30i_AOC-20sU_SamplerTeaching.pdf [In case of AOC-20i] ... Show Detail
- No:6058
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:17
-
AOCm vial sensor error[Error Code:6040 6045]
Check and perform the following. • Check whether the operating environment is appropriate. • Remove any dust or particles from the vial sensor. Show Detail
- No:6057
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 15:11
-
AOCm turret step out[Error Code:6039 6044]
Turn off the power, turn the turret gently with your hands, and check the following: • No cable is caught in the turret's notch. • No part is interfering w... Show Detail
- No:6056
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 14:53
-
AOCm L3 vial is not set[Error Code:6038 6043]
• Check whether vials conforming to the equipment are being used. • Check whether the operating environment is appropriate. Show Detail
- No:6055
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 14:52
-
AOCm L2 vial is not set[Error Code:6037 6042]
• Check whether vials conforming to the equipment are being used. • Check whether the operating environment is appropriate. Show Detail
- No:6054
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 14:51
-
AOCm solvent vial is not set[Error Code:6036 6041]
• Check whether vials conforming to the equipment are being used. • Check whether the operating environment is appropriate. Show Detail
- No:6053
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 14:50
-
AOCm waste vial is not set[Error Code:6013 6035]
• Check whether vials conforming to the equipment are being used. • Check whether the operating environment is appropriate. Show Detail
- No:6052
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 14:49
181-190 of Total 221