FAQ within [ Error/Warning ]
-
Detector is not controlled [Error Code:9021]
<GC instrument monitor setting> Press [HOME] > [Detector] on GC instrument monitor. Turn [On] the detector control and then start igni... Show Detail
- No:5943
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:43
-
Can not switch gas because of gas selector# sub error occ...
Check whether the sub gas is supplied and whether gas is not leaking from connections. After resolving the problem, press [Reset]. If the error app... Show Detail
- No:5942
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:41
-
Can not switch gas because of gas selector# main error oc...
Check whether the main gas is supplied and whether gas is not leaking from connections. After resolving the problem, press [Reset]. If the error appe... Show Detail
- No:5941
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:38
-
SCD induc. tube use time warn. [Error Code:9076]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5939
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:33
-
SCD detector abs. press is low [Error Code:9080]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5938
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:31
-
SCD flow controler#2 error [Error Code:6178]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5937
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:30
-
SCD flow controler#1 error [Error Code:6177]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5936
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:28
-
DET#1 2 3 4 failed to start plasma[Error Code:4208 4209 4...
• Check that the detector gas flow is appropriate. • Ignition may be difficult when gas replacement is insufficient immediately after tubing conne... Show Detail
- No:5935
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:17
-
System is not ready [Error Code:9010]
Start the analysis after the system is ready. If this message appears when all the parameters including temperature and flow are ready, check the followin... Show Detail
- No:5934
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 12:50
-
Pre-Run program time error [Error Code:4304]
Do not set thetotal time longer than 9999.99 min. The analysis can be performed even after this message is displayed, however, the program finishes afte... Show Detail
- No:5933
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 11:29