FAQ - Frequently Asked Questions
Keyword Search
Frequently Referenced FAQ
FAQ within [ Error/Warning ]
-
AOC link error[Error Code:4013]
Turn off the AOC power, reconnect the cable, and turn on the AOC again. If the error appears even after checking the above, turn off the main power of the in... Show Detail
- No:5994
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 09:53
-
Flow# command stopped due to error[Error Code:4021 4022 4...
When any of these errors occurs, another error should have occurred regarding the flow controller. Check the actions necessary for the error. Show Detail
- No:5995
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:01
-
DET#n TCD overcurrent detected[Error Code:4101 4102 4103 ...
Turn off the main power of the instrument, and confirm the following items. The set value of the current exceeds the maximum operating current. The mak... Show Detail
- No:5996
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:07
-
DET#n FTD overcurrent detected[Error Code:4105 4106 4107 ...
Turn off the main power of the instrument, and confirm the following items. The set value of the current exceeds the maximum operating current. The mak... Show Detail
- No:5997
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:12
-
PURn calc. prss out of range[Error Code:5042 5043 5044 50...
Set the purge flow within the AFC control range. Show Detail
- No:5998
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:14
-
CARm calc. ratio out of range[Error Code:5075 5076 5108 5...
Change the split ratio so that the following equation is satisfied. (Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow) Show Detail
- No:5999
- Date of Issue:2021/05/18 15:36
- Last Updated:2023/12/13 13:23
-
Analysis waiting[Error Code:5112]
Set [Prep Run Wait Time] longer. The PREP RUN is activated when using the following functions. Splitless (split in standby) Carrier gas saver (saved... Show Detail
- No:6000
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:19
-
CARm AFC set point not reached[Error Code:2130 2131 2204 ...
Check that the gas main valve is open. Check that the set values of the column inlet pressure and total flow are within the AFC control range. Check that... Show Detail
- No:6001
- Date of Issue:2021/05/18 15:36
- Last Updated:2023/12/13 13:22
-
CARn was overpressure[Error Code:2214 2215 2216 2217]
Do not open or close the main valve of the supply gas during AFC operation. Restart the GC with the gas supplied. Check that the gas supply pressure is w... Show Detail
- No:6002
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:40
-
PURn was overpressure[Error Code:2218 2219 2220 2221]
Check that the purge line is not blinded. Do not open or close the main valve of the supply gas during AFC operation. Restart the GC with the gas supplie... Show Detail
- No:6003
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/07 17:03
101-110 of Total 221