FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030
Keyword Search
GC-2030
Frequently Referenced FAQ
FAQ within [ GC-2030 ]
-
DetAPCn makeup set point not reached[Error Code:2014 2017...
・Check that the gas main valve is open. ・Check that the gas supply pressure is within the range below. ・Check that the supply pressure is sufficien... Show Detail
- No:5966
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:26
- Category: Error/Warning
-
Oven door is open[Error Code:1061]
Close column oven door and press [Reset]. If the error appears even with the oven door closed, turn off the main power of the instrument, stop gas supply, a... Show Detail
- No:5965
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:19
- Category: Error/Warning
-
DET#n set temperature is low[Error Code:1049 1050 1051 1052]
・To prevent detector contamination, set a detector temperature higher than the column oven temperature. ・When setting a detector temperature lower than the ... Show Detail
- No:5964
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 10:26
- Category: Error/Warning
-
COL CRG is out of control[Error Code:1034]
・When oven door opens, close the door and press [Reset]. ・When coolant becomes empty, turn off the main power of the instrument and replace coolant. ・If th... Show Detail
- No:5963
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 10:11
- Category: Error/Warning
-
Heat is escaping[Error Code:1001]
・When oven door opens, close the door and press [Reset]. ・When the insulation of the oven has a hole, turn off the main power of the instrument and close th... Show Detail
- No:5962
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/11 15:01
- Category: Error/Warning
-
Prep run wait time elapsed[Error Code:9062]
For manual injection analyses, introduce the sample within the [Prep RUN Wait Time]. For analyses with extended pre-run time, such as for HS warming, set [P... Show Detail
- No:5961
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 09:59
- Category: Error/Warning
-
Gas Selector# sub was overpressure[Error Code:2368 2369]
Check the supply gas pressure. If the supply pressure is correct, turn off the main power of the instrument, stop gas supply, and contact your S... Show Detail
- No:5960
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 09:51
- Category: Error/Warning
-
Gas Selector# purge was overpressure[Error Code:2366 2367]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5959
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 09:48
- Category: Error/Warning
-
Gas Selector# main was overpressure[Error Code:2364 2365]
Check the supply gas pressure. If the supply pressure is correct, turn off the main power of the instrument, stop gas supply, and contact your S... Show Detail
- No:5958
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 17:30
- Category: Error/Warning
-
SCD unable to auto start[Error Code:9078]
If a leak check has failed,check the following and perform a leak checkagain. Screen_SCD-Leak-Check.pdf Cause Solution ... Show Detail
- No:5957
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 17:21
- Category: Error/Warning
31-40 of Total 225