SCD furnace temperature set point not reached [Error Code...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5854
- Date of Issue:2021/05/18 15:38
- Last Updated:2021/06/04 11:42
- Category: Error/Warning
-
SCD furnace temperature controller error [Error Code:6105]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5853
- Date of Issue:2021/05/18 15:38
- Last Updated:2021/06/04 11:40
- Category: Error/Warning
-
SCD furnace temperature exceeds the limit [Error Code:6104]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5852
- Date of Issue:2021/05/18 15:38
- Last Updated:2021/06/04 11:38
- Category: Error/Warning
-
SCD furnace temperature sensor error [Error Code:6103]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5851
- Date of Issue:2021/05/18 15:38
- Last Updated:2021/06/04 11:36
- Category: Error/Warning
-
SCD furnace temperature sensor is down [Error Code:6101]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5850
- Date of Issue:2021/05/18 15:38
- Last Updated:2021/06/04 09:55
- Category: Error/Warning
-
DET#1 2 3 4 flame is out[Error Code:4109 4110 4111 4112]
Check that the column is connected. Check also that there is no leak at column connections. Check that the column insertion length is appropriate. Ca... Show Detail
- No:5849
- Date of Issue:2021/05/18 15:38
- Category: Error/Warning
- Brached FAQ
-
SCD hydrogen was overpressure[Error Code:6164]
・Check that the gas supply pressure is within the range below. ・If the gas supply pressure is appropriate, turn off the main power of the in... Show Detail
- No:5970
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:49
- Category: Error/Warning
-
APCy set point not reached[Error Code:2026 2027 2028 2029...
・Check that the gas main valve is open. ・Check that the gas supply pressure is higher than the set pressure. ・Check that no tubing connections are loose an... Show Detail
- No:5969
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:44
- Category: Error/Warning
-
DetAPCn air set point not reached[Error Code:2016 2019 20...
・Check that the gas main valve is open. ・Check that the gas supply pressure is within the range below. ・Check that the supply pressure is sufficient ... Show Detail
- No:5968
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:39
- Category: Error/Warning
-
DetAPCn hydrogen set point not reached[Error Code:2015 20...
・Check that the gas main valve is open. ・Check that the gas supply pressure is within the range below. ・Check that the supply pressure is sufficient ... Show Detail
- No:5967
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 12:32
- Category: Error/Warning