1. X-rays will be generated. Close the sample chamber cover. 2. Double click theor icon on the Desktop to launch the software. 3. Initialize th... Show Detail
- No:5612
- Date of Issue:2021/04/06 14:38
- Brached FAQ
-
(EDX) Changing and Supplying Oil to the Rotary Pump (When...
The rotary pump used as an auxiliary pump for the evacuation system requires oil change every one and a halfyears*. Without oil change, insufficient vacuum, ... Show Detail
- No:5614
- Date of Issue:2021/04/07 10:21
-
(EDX) The instrument is not turned ON.
Check Points1:Check that the power cable isconnected correctly. Remedies1:Connect the power cable correctly. Check Points2:Check that power is su... Show Detail
- No:5615
- Date of Issue:2021/04/07 10:28
-
(EDX) Analysis cannot be started. "Unknown" or "Cannot ge...
Check Points1:Check that the power switch of thespectrometer is ON. Remedies1:Turn the power to the spectrometer ONand restart the software. Check ... Show Detail
- No:5616
- Date of Issue:2021/04/07 10:30
-
(EDX) No spectrum name is identified. The intensity indic...
Check Points:Use the sample for energy calibration,and check if peaks appear at thecorrect positions. Remedies:Perform energy calibration.Perform an energ... Show Detail
- No:5617
- Date of Issue:2021/04/07 10:31
-
(EDX) An abnormal sound or smell is being produced from t...
Turn the power OFF and unplug theinstrument. Contact a Shimadzu sales or service representative. Show Detail
- No:5618
- Date of Issue:2021/04/07 10:32
-
(EDX) The sample chamber cover does not open smoothly or ...
Adjustment or part replacement isnecessary. Contact a Shimadzu sales or service representative. Show Detail
- No:5619
- Date of Issue:2021/04/07 10:33
-
(EDX) The sample has fallen off the beam window into the ...
Removal of the sample, adjustmentand verification of the inside of thespectrometer are necessary. Contacta Shimadzu sales or service representative Show Detail
- No:5620
- Date of Issue:2021/04/07 10:34
-
(EDX) The sample image is faint.
Remedies1: Adjust the light level in the PCEDXNaviwindow. (EDX-LE):Adjust the light level in the [ScreeningAnalysis] window. Remedies2: The sample... Show Detail
- No:5621
- Date of Issue:2021/04/07 10:36
-
(EDX) Quantitative calculation of FP method
With the X-ray fluorescence spectrometry, if the composition (elements and content) of the sample is known, X-ray fluorescence intensities can be calculated ... Show Detail
- No:5665
- Date of Issue:2021/04/28 13:18
- Last Updated:2021/05/20 08:34