(EDX) No spectrum name is identified. The intensity indic...
Check Points:Use the sample for energy calibration,and check if peaks appear at thecorrect positions. Remedies:Perform energy calibration.Perform an energ... Show Detail
- No:5617
- Date of Issue:2021/04/07 10:31
-
(EDX) Analysis cannot be started. "Unknown" or "Cannot ge...
Check Points1:Check that the power switch of thespectrometer is ON. Remedies1:Turn the power to the spectrometer ONand restart the software. Check ... Show Detail
- No:5616
- Date of Issue:2021/04/07 10:30
-
(EDX) The instrument is not turned ON.
Check Points1:Check that the power cable isconnected correctly. Remedies1:Connect the power cable correctly. Check Points2:Check that power is su... Show Detail
- No:5615
- Date of Issue:2021/04/07 10:28
-
(EDX) Changing and Supplying Oil to the Rotary Pump (When...
The rotary pump used as an auxiliary pump for the evacuation system requires oil change every one and a halfyears*. Without oil change, insufficient vacuum, ... Show Detail
- No:5614
- Date of Issue:2021/04/07 10:21
-
1. X-rays will be generated. Close the sample chamber cover. 2. Double click theor icon on the Desktop to launch the software. 3. Initialize th... Show Detail
- No:5612
- Date of Issue:2021/04/06 14:38