• Font Size
  • S
  • M
  • L

FAQ within [ Energy Dispersive X-ray Fluorescence Spectroscopy(EDX) ]

31-35 of Total 35

4 / 4page
  • (EDX) Unit of "%" in "Easy" analysis

    The unit of the values obtained in easy analysis mode is wt%. Quantitative values can also be calculated in atom%. In quantitative mode, this can be set pr... Show Detail

    • No:5708
    • Date of Issue:2021/04/28 13:55
    • Last Updated:2021/05/20 09:04
  • (EDX) Analytical depth

    The analysis depth for fluorescent X-rays changes with the principal components of the sample and the target elements (analysis lines). •The lighter the ele... Show Detail

    • No:5710
    • Date of Issue:2021/04/28 13:57
    • Last Updated:2021/05/20 09:05
  • (EDX) High-concentration "Sn" aqueous solution analysis

    When measuring high-concentration aqueous solutionswith fluorescent X-rays, the fluorescent X-ray intensity may fail to risedue to the self-absorption phenom... Show Detail

    • No:5712
    • Date of Issue:2021/04/28 13:58
    • Last Updated:2021/05/20 09:06
  • (EDX) Thread plating film thickness measurement

    Plating thickness can be measured by the EDX. There are two quantitative calculation methods, the calibration curve method and the thin-film FP method. For... Show Detail

    • No:5714
    • Date of Issue:2021/04/28 13:59
    • Last Updated:2021/05/20 09:07
  • (EDX) "Pb" measurement in solder

    This is likely due to sample segregation. The Pb in the solder is sometimes segregated at the surface. EDX analysis is strongly affected by the sample surf... Show Detail

    • No:5716
    • Date of Issue:2021/04/28 14:01
    • Last Updated:2021/05/20 09:07

31-35 of Total 35

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.