FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis > Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
Keyword Search
Frequently Referenced FAQ
FAQ within [ Energy Dispersive X-ray Fluorescence Spectroscopy(EDX) ]
-
(EDX) "Pb" measurement in solder
This is likely due to sample segregation. The Pb in the solder is sometimes segregated at the surface. EDX analysis is strongly affected by the sample surf... Show Detail
- No:5716
- Date of Issue:2021/04/28 14:01
- Last Updated:2021/05/20 09:07
-
(EDX) Thread plating film thickness measurement
Plating thickness can be measured by the EDX. There are two quantitative calculation methods, the calibration curve method and the thin-film FP method. For... Show Detail
- No:5714
- Date of Issue:2021/04/28 13:59
- Last Updated:2021/05/20 09:07
-
(EDX) High-concentration "Sn" aqueous solution analysis
When measuring high-concentration aqueous solutionswith fluorescent X-rays, the fluorescent X-ray intensity may fail to risedue to the self-absorption phenom... Show Detail
- No:5712
- Date of Issue:2021/04/28 13:58
- Last Updated:2021/05/20 09:06
-
The analysis depth for fluorescent X-rays changes with the principal components of the sample and the target elements (analysis lines). •The lighter the ele... Show Detail
- No:5710
- Date of Issue:2021/04/28 13:57
- Last Updated:2021/05/20 09:05
-
(EDX) Unit of "%" in "Easy" analysis
The unit of the values obtained in easy analysis mode is wt%. Quantitative values can also be calculated in atom%. In quantitative mode, this can be set pr... Show Detail
- No:5708
- Date of Issue:2021/04/28 13:55
- Last Updated:2021/05/20 09:04
-
• Strongly magnetic samples cannot be measured. •Measurement is possible if the magnetism at a distance of 7cm from the sample is 20 gauss or less. •If pos... Show Detail
- No:5706
- Date of Issue:2021/04/28 13:54
- Last Updated:2022/09/20 13:58
-
(EDX) "Hg" detected by RoHS analysis in resin containing ...
•Hg was detected due to the influence of the BrKαESC peak. →HgLα: 9.99 keV, BrKαESC: 10.16 keV •With the EDX, when high intensity peaks are observed, it i... Show Detail
- No:5704
- Date of Issue:2021/04/28 13:52
- Last Updated:2021/05/20 09:02
-
(EDX) The quantitative values of PbLα and PbLβ1 are signi...
Look at the profile, and check the peaks. •Confirm the PbLα(10.55keV) and PbLβ1(12.62keV) peak intensity ratio. → PbLα : PbLβ1 ≒ 1 : 1 (normal) •If PbLα ... Show Detail
- No:5701
- Date of Issue:2021/04/28 13:50
- Last Updated:2021/05/20 08:59
-
(EDX) Negative quantitative value display in quantitative...
At 0ppm, the variance in X-ray intensity will be a negative value if it is below the average intensity. A negative value is treated as a quantitative value ... Show Detail
- No:5699
- Date of Issue:2021/04/28 13:47
- Last Updated:2021/05/20 08:58
-
(EDX) Plating film thickness measurement (multiple layers)
• There is no way to specify what layer generated the fluorescent X-rays, and so analysis is not possible if all of the layers are unknown. •If there is 1 u... Show Detail
- No:5697
- Date of Issue:2021/04/28 13:46
- Last Updated:2021/05/20 08:57
1-10 of Total 35