FAQ within [ FTIR Spectroscopy(FT-IR) ]
-
• The IRAffinity-1 series has a self-diagnosis function. The IRAffinity-1 series automatically checks its own status during operation or startup. When all it... Show Detail
- No:4905
- Date of Issue:2021/01/27 10:38
- Last Updated:2022/06/28 13:29
- Brached FAQ
-
(FTIR) Installation of Sample Compartment Window
It needs to be replaced by our service personnel. Please contact a Shimadzu sales or service representative. 1. Remove the Sample compartment window hold... Show Detail
- No:4910
- Date of Issue:2021/01/27 10:38
- Last Updated:2022/06/28 13:51
- Brached FAQ
-
(FTIR) The signal intensity of the power spectrum remains...
Reason 1. The light source is broken. 2. The fuse has blown. 3. Selection of the detector is wrong. 4. Something that intercepts the light path is in... Show Detail
- No:4906
- Date of Issue:2021/01/27 20:30
-
(FTIR) A Part of the Mirror Is a Little White
Reason It seems to be the result of the heat of the infrared light source for an extended time. Solution Conduct the performance check according t... Show Detail
- No:4907
- Date of Issue:2021/01/27 20:30
-
(FTIR) Power Spectrum Becomes too Small, or Fluctuates
Reason 1. Deterioration because of long term use 2. Trouble with the power board 3. Trouble with the detector Solution 1. Deterioration becaus... Show Detail
- No:4908
- Date of Issue:2021/01/27 20:30
-
(FTIR) IRSpirit Wavenumber Calibration
Execute Wavenumber calibration of IRSpirit using peaks of carbon dioxide and water vapor. FTIR spectrum is described with wavenumber unit. Since wavenumber ... Show Detail
- No:4909
- Date of Issue:2021/01/27 20:50
-
(FTIR) IRTracer-100 Replacement of Beam Splitter Unit
The scan wavenumber range of the IRTracer-100 can be changed by switching the beam splitter unit. If it is expected that the main power supply will be off fo... Show Detail
- No:4911
- Date of Issue:2021/01/27 21:25
-
(FTIR) Suitable Locations and Preparation
Operation guarantee temperature range:10 to 30ºC, within the temperature fluctuation range ±2 ºC/h Installationtemperature range: • IRSpirit / IRXross:... Show Detail
- No:4912
- Date of Issue:2021/01/27 21:40
- Last Updated:2022/06/28 13:44
-
(FTIR) FTIR Analysis Basic Qestion
Click here forFTIR Analysis Basic Qestion. The following Q & A is described. Q: Is FTIR affected by the environment? Q: How deep does the infrared... Show Detail
- No:4913
- Date of Issue:2021/02/02 08:54
-
(FTIR) Noise Appears at High Absorbance Peak Top
When performing a Fourier transform on a finite interval of data,computational noise is likely to occur due to discontinuities in thedata at both ends, and t... Show Detail
- No:8876
- Date of Issue:2022/06/28 15:36