FAQ - Frequently Asked Questions
- Products Search > GC-MS > Troubleshooting
Keyword Search
Frequently Referenced FAQ
FAQ within [ Troubleshooting ]
-
(GCMS) Chemical Ionization(CI, NCI) Excessive Introductio...
Sometimes too much reagent gas is introduced during reagent gas adjustment. If this happens, extract excess reagent gas by following the procedure below. ... Show Detail
- No:4689
- Date of Issue:2020/12/17 10:16
-
(GCMS) Chemical Ionization(CI, NCI) When good spectrum ca...
There may be leak of the reagent gas if the conditionsare not satisfied in spite that the reagent gas is being supplied. The possible causes of leak are desc... Show Detail
- No:4702
- Date of Issue:2020/12/17 12:59
-
(GCMS) Chemical Ionization(CI, NCI) When signal intensit...
Cause 1: Either the cylinder valve of the gas cylinder, the control knob of the pressure regulator, or the outlet valve of the pressure regulator is closed. ... Show Detail
- No:4703
- Date of Issue:2020/12/17 12:59
-
(GCMS) Measurres durriing Powerr Outtage
For power outages of 10 minutes or less, normally no particular procedure is required as long as the [Vacuum Restart Mode] checkbox is selected in the window... Show Detail
- No:4704
- Date of Issue:2020/12/17 13:21
-
(GCMS) Error status indicated by LED on MS
Normal LED Indications: Switch MS Unit → All 3 LED light → Green and Red LED goes out → Yellow LED remains ON LED Error Indications Yellow LED... Show Detail
- No:4706
- Date of Issue:2020/12/17 15:49
-
(GCMS) GCMSsolution Software Startup Errors
1. PC and MS or MS and GC connected properly? All connections must be made properly 2. Is GC power on? Turn on GC power supply and restart. 3. Is MS po... Show Detail
- No:4707
- Date of Issue:2020/12/17 15:54
-
(GCMS) Vacuum System Auto Startup Errors
Cause 1: Does Rotary Pump1 (RP1) operate? Action 1: Connect the RP1 power cable to the back panel of the MS, and turn on the RP1 switch. If not resolve... Show Detail
- No:4708
- Date of Issue:2020/12/17 16:03
-
• "Filament consumed" message appears. Filament installation is not correctly. Filament malfunction. Click here for Filament Replacement Procedur... Show Detail
- No:4709
- Date of Issue:2020/12/17 16:18
-
(GCMS) Autotuning and Analysis Errors
1. Error message displayed when standard is introduced. Checks that Filament ON Errors. 2. "Initialization Error" message appears. Whenstandar... Show Detail
- No:4710
- Date of Issue:2020/12/17 16:28
-
(GCMS) No peaks appear, or peaks are too small
Cause 1: Carrier gas flows? Action 1: If carrier gas is not flowing, please let the carrier gas flow. Cause 2: Carrier gas leaks. Action 2: Tigh... Show Detail
- No:4711
- Date of Issue:2020/12/17 16:38
1-10 of Total 12