FAQ - Frequently Asked Questions
- Products Search > GC-MS > Troubleshooting
Keyword Search
Frequently Referenced FAQ
FAQ within [ Troubleshooting ]
-
(GCMS) "Detector saturated" message appears during analysis
1. Turn on the filament after solvent has completely eluted. If peaks elute before the solvent, use a time program to turn the filament off and on. 2. Dec... Show Detail
- No:4712
- Date of Issue:2020/12/17 16:43
-
(GCMS) RF Temperature Control Error
1. The room temperature might be hight (such as over 40ºC). Check environmental conditions at the installation site. 2. The Q1 or Q3 RF temperature contro... Show Detail
- No:4713
- Date of Issue:2020/12/17 16:46
11-12 of Total 12
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.